AT2000
 

A/T 2000M

The A/T 2000M was manufactured between 1985 - 1998.
Limited parts available.

Specifications

Electrical: 115 V, 60 Hz, 8 amps. Use a 15 amp. 3 prong, grounded outlet.
Water Flow: 1/2 gallon per minute water flow (while processing film, a water softener should not be used).
Water Pressure: 80 PSI maximum/30 PSI minimum. Water source must have a manual shut-off near processor.
Water Connection: Valved 3/4" male straight pipe fitting at rear of processor.
4 ft. tubing supplied: 3/4" female straight pipe fitting x female garden hose fitting. Connect to 3/4" male garden hose fitting with shutoff..
Water Backflow Prevention: A vacuum breaker is not normally required. The processor has a 1" air gap between water supply inlet and maximum water level height.
Water Temperature: Cold water supply (water temp should not drop below 45° F or rise above 100° F).
Drain: Open, vented PVC trap; bo copper or brass should be used.
Dimensions: A/T 2000M: 16" (H) x 15" (W) x 25" (D); Automatic Replenisher: 17" (H) x 16" (W) x 14" (D)
Weight: 92 lbs. with water and chemistry (77 lbs. empty)
Ventilation: Darkroom temperature must be below 85°F (29.5°C) during processing. Because heat is generated during processing, adequate ventilation is required to maintain darkroom temperature.