ALLPRO 2010 Film Processor
Isn't it time to take a look at the latest film processing system from ALLPRO Imaging?
The ALLPRO 2010® is the next generation of film processors from ALLPRO Imaging. Like its predecessor, the 2000M, the ALLPRO 2010 consistently delivers high quality 8"x10" and 10"x12" x-rays quickly and automatically.
The ALLPRO 2010's microprocessor takes the guesswork out of film processing by accurately controlling time, temperature and chemistry strength, the keys to consistent film quality. Two user-selectable speeds allow the processing of all types of x-ray films. The efficient warm air dryer ensures that films are delivered completely dry to allow immediate reading and diagnosis.
- Consistent reliable film quality
- Fast processing time
- Easy cleaning and maintenance
- Space-saving convenience
Models
| Part Number |
Description |
| 45002 |
All-Pro 2010 Automatic Film Processor for film up to 10 inches wide |
| 45002W |
All-Pro 2010 Automatic Film Processor with water recirculation |
Accessories
| Daylight Loader |
P/N 45125 - Dimensions (W x D x H): 16 ¼" x 13 ¾" x 10 ¾" |
Quickloader™ Covered Feed Chute |
P/N 45785 - Dimensions (W x D x H): 6" x 15 ¾" x 12 3/8" |
Formula 2000 Twin Pack |
P/N 43945 2 bottles per case |
| Spray 2000 |
P/N 43965 - 1 bottle per case |
Chemistry Recovery Containers |
P/N 44160: 5 gallons, 2 each
P/N 44170: 2 ½ gallons, 2 each
|
Specifications
| Electrical: |
115 V, 60 Hz, 10 amps. Use a 15 amp. 3 prong, grounded outlet.
A separate dedicated 15 AMP line is recommended. |
| Water Flow: |
1/2 gallon per minute water flow (while processing film). |
| Water Pressure: |
80 PSI maximum/30 PSI minimum. Water source must have a manual shut-off near processor. |
| Water Connection: |
Valved 3/4" male garden hose fitting near rear of processor. 6' hose with standard female garden hose fittings provided. |
| Water Backflow Prevention: |
A vacuum breaker is not normally required. The processor has a 1" air gap between water supply inlet and maximum water level height. |
| Water Temperature: |
55°F (13°C) to 85°F (29.5°C). |
| Drain: |
1 1/2" vertical PVC standpipe, open at top. DO NOT DRAIN INTO COPPER OR BRASS. The top of the standpipe must be 6" below the bottom of the countertop. |
| Dimensions: |
18" (H) x 15" (W) x 25" (D) with leveling base. |
| Weight: |
90 lbs. with water and chemistry (75 lbs. empty) |
| Ventilation: |
Room temperature must be below 85°F (29.5°C) during processing. Because heat is generated during processing, adequate ventilation is required to maintain darkroom temperature. 8-12 volume room air changes per hour recommended. |

Marketing Literature